Optical sensor tracks radiotherapy treatment site breast motion

2009 01 15 16 03 49 510 Med Phys Web Logo 100 Postsurgical radiotherapy is an effective way of preventing breast tumors from reappearing close to the treatment site. Applying this strategy reduces the incidence of local recurrence by two-thirds. The downside is the mismatch between ideal treatment plans and the doses actually delivered across the breast. This variance between theory and practice can be particularly significant in large-breasted women, raising the risk of treatment failure and/or damage to healthy tissue.

Although conformal methods of delivery could address this issue, switching delivery techniques raises another potential problem: Will breast motion compromise the accuracy of a highly targeted radiotherapy regimen?

The considered verdict at the moment is: "We're not sure," which isn't really good enough. Patients and practitioners need a more definitive answer if methods such as intensity-modulated radiation therapy (IMRT) are to be used routinely.

A recent paper in Physics in Medicine and Biology (November 7, 2009, Vol. 54:21, pp. 6515-6534) shows how this information could be acquired using an optical sensor system. The system relies on a 3D method of optical measurement known as Fourier profilometry, in which a fringe pattern is projected onto the patient. The 3D shape of the skin surface is then recovered from the modulation of the fringes.

The optical technique has many advantages over alternative methods, said Gareth Price, a researcher in the radiotherapy physics department at the Christie NHS Foundation Trust (Manchester, U.K.) and lead author of the paper. Measurements are acquired in real-time and without exposing patients to any additional ionizing radiation. "We can analyze the surface at a typical video frequency, which is about 25 Hz. It's ideal for measuring dynamic motion, such as breathing," he told medicalphysicsweb.

The sensor system was used on 13 patients at Christie scheduled for postsurgical IMRT for breast cancer. Study subjects were divided into two groups based on breast volume for the purpose of data analysis, but all underwent the same procedure.

Data were acquired during treatment-beam delivery for an average of 12 fractions per patient. Video footage lasting 20 s (500 sequential surfaces) was analyzed for each patient fraction. This equated to approximately five breathing cycles. The results of the analysis were displayed as a set of three surface maps for each fraction -- one looking at the dominant frequency of motion, another focusing on motion amplitude, and the third representing the lag of each surface point with respect to a fixed marker.

In general, breast motion was found to be reasonably stable between and during treatment fractions. The motion frequency was found to be similar for both patient groups (3-5 s), though significant intrafraction variability (1-2 s) was observed in the small-breasted cohort. The mean peak-to-trough amplitude was also similar in both groups (2-4 mm).

Researchers were also able to see how the breast deformed during motion, with the inside of the sternum tending to lead the breast volume. The phase difference was greater in larger-breasted patients (about 0.4 s) than in the smaller-breasted group (about 0.3 s) as might be expected.

So does this motion make any difference to dose delivery? That's the next question to be addressed, Price said. "We can take the surface from the planning scan, change it slightly using the results we have got from this paper, do another plan and see how the dose distribution changes," he said. "We are just starting to look at this."

By Paula Gould
Medicalphysicsweb contributing editor
November 26, 2009